![AND Weighing Indicator AD-4401 Скачать руководство пользователя страница 53](http://html1.mh-extra.com/html/and/weighing-indicator-ad-4401/weighing-indicator-ad-4401_instruction-manual_2941326053.webp)
51
7-6 COMPARISON WEIGHT
Although the weight compared with the setpoint is mainly the net, the gross is partially compared. Since the net
assumes a negative value in loss-in-weight, setpoint is made with the polarity reversed.
Setpoint of the weight with the setpoint allows you to select the display count and internal count.
One graduation of the weight value appearing in the display section is the minimum graduation set with CALF-03.
Provided internally is the weight value called high-resolution “internal count” with the minimum graduation calculated as 1.
For example, when the minimum graduation is 10, a display changes only in a step of 10, but a finer change is captured
internally.
Batch weighing mode mainly uses the internal count, and Check comparison mode uses the display count.
This is because the internal count is superior from a viewpoint of weighing accuracy, but if the platform scale is compared
using the internal count, it appears as if the displayed weight and comparison result are contradictory to each other.
The comparison weight is selected with the general function, “Selection of Setpoint Weight SQF-01.” It is meaningless to
use the internal count, unless the minimum graduation is 2 or more.
7-7 AUTO PRINT
Auto print is the capability to output the Batch Finish weight from the external output where the output mode has been set
to Auto Print. The output timing is as follows depending on the weighing mode.
CUSTOMER PROGRAMMED CONTROL MODE (NORMAL BATCHING , LOSS-IN-WEIGHT)
Since the weighing completion timing is not definite in customer programmed control mode, the weight is output only
once when a setpoint target net value is stabilized at 5D or greater.
To output again, the setpoint target net value must be to less than 5D (5 graduations).
BUILT-IN-AUTOMATIC PROGRAM MODE
The weight is output the moment the weighing sequence encounters the batch finish timing.
7-8 AUTOMATIC ACCUMULATION
The Automatic accumulation function automatically accumulates the net weight after weighing. The its timing is the same
as auto print, but its functioning differs depending on “Automatic accumulation SQF-02.”
Setting
SQF- Name
Default
Parameter Description
0 No
Automatic
accumulation
1
Accumulates only acceptable weight
02
Automatic
accumulation
0
2
Accumulates all values
Содержание Weighing Indicator AD-4401
Страница 1: ...AD 4401 WEIGHING INDICATOR 1386 2A IE...
Страница 42: ...40...
Страница 87: ...85 APPENDIX DIMENSIONS AD 4401 OP 05...
Страница 88: ...86...
Страница 89: ...87...
Страница 90: ...88...