![AND Multi Function Weighing Indicator AD-4402 Скачать руководство пользователя страница 23](http://html.mh-extra.com/html/and/multi-function-weighing-indicator-ad-4402/multi-function-weighing-indicator-ad-4402_instruction-manual_2941316023.webp)
AD-4402 Page
21
5.1.
5.1.
5.1.
5.1.
Actual Load Calibration (using
Actual Load Calibration (using
Actual Load Calibration (using
Actual Load Calibration (using a
a
a
a Mass)
Mass)
Mass)
Mass)
ESC
key
If you want to return to the weighing mode during the
calibration mode, press the
ESC
key anytime. It is
effective until the last displayed parameter.
Example: zero adjustment only, etc.
ENTER
key When the key is pressed, the procedure stores the
current parameter and proceeds to the next step.
Step 1 Press and hold the
ENTER
key and press the key to display the
menu in the weighing mode.
Step 2 Press the key twice to select the menu
CAL
.
Press the
ENTER
key to enter the calibration mode.
Step 3 Press the
ENTER
key to enter the menu
CAL
.
Step 4 Select a unit using the numerical keys and press the
ENTER
key
to store it.
Step 5 Select a decimal point using the numerical keys and press the
ENTER
key to store it.
Step 6 Select a minimum division using the numerical keys and press the
ENTER
key to store it.
Step 7 Select a weighing capcity using the numerical keys and press the
ENTER
key to store it.
Step 8 Perform the zero point adjustment.
Place nothing on the weighing pan and press the
ENTER
key to
store it after the
STABLE
indicator is displayed.
Whether
the
STABLE
indicator is displayed or not, if you want to
store it, wait for ten seconds and press the
ENTER
key.
Step 9 If the
F1
key is pressed, it will branch out to the digital span.
Step 10 Specify a total mass value to place on the weighing pan using the
numerical keys and press the
ENTER
key to store it.
Step 11 Place the specifyed mass on the weighing pan and press the
ENTER
key to store it after the
STABLE
indicator is displayed.
Whether
the
STABLE
indicator is displayed or not, if you want to
store it, wait for ten seconds and press the
ENTER
key.
Step 12 Press the
ESC
key to return the weighing mode.