![AND Multi Function Weighing Indicator AD-4402 Скачать руководство пользователя страница 104](http://html.mh-extra.com/html/and/multi-function-weighing-indicator-ad-4402/multi-function-weighing-indicator-ad-4402_instruction-manual_2941316104.webp)
Page
102
AD-4402
9.4.
9.4.
9.4.
9.4.
Initializing Parameters
Initializing Parameters
Initializing Parameters
Initializing Parameters
This function initializes the parameters stored in the indicator.
The parameters are stored in the flash memory and backup RAM.
Caution
There are reset functions that require re-calibration of the indicator
Note where the parameters are stored.
Kinds of intialization mode
Kinds of intialization
Description
Initializing RAM
The backup RAM is reset. Zero point of the gross display,
tare value zeroes.
Initializing material code or
recipe code
Material code and recipe code is reset.
Initializing the function list
Resets parameters of the function list in flash memory.
Initializing caribration data
Resets parameters of calibration data in flash memory. If
this function is used, calibrate the indicator.
Initializing all parameters
Resets all parameters, calibrate the indicator.s
The location of the parameters and objects of initialization mode
Material
code
Recipe
code
Zero
Tare
Ma
te
ria
l co
de
Total d
a
ta
R
e
ci
pe co
de
Total d
a
ta
Function
list
C
a
lib
ra
ti
on dat
a
Backup RAM
o
o
o
o
o
o
Location
Flash memory
o o o
o
Initializing RAM
o
o
Initializing material code or
recipe code
o
o
o
o
Initializing the function list
o
Initializing
caribration
data o
In
itia
liza
tion
mo
de
Initializing all parameters
o
o
o
o
o
o
o
o