19
User manual Zitzi Starfish Pro
www.anatomicsitt.com | +46 11 16 18 00
Buzzer functionality
The buzzer will make a warning when a button on the hand control is pressed and
the battery capacity is low.
The buzzer can also be activated by the control box to signal other conditions. This
must be specified in the control box software.
Deep discharge protection
•
The battery BA21 has a protection against deep discharge to protect the battery
life. The protection is activated when the battery is discharged.
•
Charge the battery to exit the deep discharge mode. Ensure that the battery is
sufficiently charged before use.
If the battery is completely discharged, the charging will be started at a very small
rate to protect the battery. In this case the yellow LED will be flashing. If the battery
does not stop flashing and start charging normally within 12 hours (LED – solid
yellow), the battery is defect and must be replaced.
Follow the disposal instructions.
Safety feature
The battery contains several mechanisms to protect itself from being damaged due
to excessive use.
In case of overheating, the device will activate a thermal protection. No power
output will be available until the temperature has returned to normal operating
range. Overheating can occur through extensive use, at high temperatures or
through overheating.
Duty cycle guideline: approx. 5% (1 minute continuous use followed by 19 minutes
not in use).
Содержание Zitzi Starfish Pro
Страница 2: ...2 User manual Zitzi Starfish Pro www anatomicsitt com 46 11 16 18 00 ...
Страница 47: ......