Page 8
Probe hose
⇒
Connect the probe hose to the GAS connection.
If not already fixed, position the water separator.
Water separator
⇒
Close the measuring probe with the red sealing cap provided. Then press the key
and the
hydrostatic test starts which counts down from 40.
This appears on the display:
Hydrostatic test
40
In the event of a leak, the following appears on the display:
Leaky probe or
water separator
In this case, you must check the device for any leaks.
Possible sources of faults: condensate vessel, seals in condensate vessel, various hose connections, front
part of probe or filter. Once you have located the source of the fault, press the
key and the
hydrostatic test will be started again.
If it is impossible to achieve a leak-free device, it must be checked over by anapol.
As soon as the device is sealed, this appears on the display:
Leak tightness OK
Probe to
fresh air
⇒
Remove the red sealing cap and press the
key: the pump starts up and the device begins the
calibration process.
2.2
Calibration
The calibration procedure starts now.
Position the fresh air probe where the combustion plant to be measured draws
in its fresh air.
The device is now in the calibration phase. The following appears on the display: