SPECORD
®
Accessories
Standard cell holders
7 Accessories
7.1
Standard cell holders
The standard equipment of the SPECORD
®
includes two cell holders for cells of up to 50
mm pathlength.
The mount for the cell holder is located on the right-hand sample compartment wall in
front of the beam-exit window (
→
Fig. 4-5 Sample compartment, p. 17). The cell holder
is designed for cells having a pathlength of up to 50 mm and a width of 12.5 mm.
1 Cell
support
2 Guide
Fig. 7-1
Standard cell holder
Push the cell holder with its guide (2) from top onto the cell-holder mount.
The beam transmits the cells at a height of approximately 5 to 15 mm above the support
of the cell holder.
Inserting the cell
1. Fill the sample into the cell up to a level of at least 20 mm.
2. Insert the cell so it rests against the right-hand support surface of the cell holder.
Correct
Fig. 7-2
Inserting the cell
SPECORD
®
200/205/210/250 User's Manual – Edition 04/2005
Page 27
Analytik Jena AG