Installation and start-up
SPECORD
®
Connecting the SPECORD
3. Using a screwdriver, turn the transport lock screw
(1)
clockwise as far as it will go.
6.3
Connecting the SPECORD
Observe correct position of device!
Move the SPECORD
®
only in upright position (observe label on transport case).
Do not turn the device upside down.
Release transport lock!
Loosen the transport lock screw located on the rear left of the SPECORD
®
.
Electric shock
Before connecting the SPECORD
®
and the PC to line power, make sure the available
line voltage agrees with the operating voltage specified on the type labels. Operation
of the devices on line voltages other than the specified operating voltage may result in
their destruction.
Standard equipment
The standard equipment of the SPECORD
®
includes the following items:
•
2 cell holders
•
Installation CD-ROM of WinASPECT
®
basic module
•
Spare halogen lamp
•
Deuterium lamp
•
Power cable
•
PC – SPECORD
®
connecting cable
•
Spare fuses 5 x 20 T 3.15 A L
•
User's Manuals for the SPECORD
®
, the accessories and the WinASPECT
®
control and analysis software.
Connecting SPECORD
®
and PC
1. Remove the device from its transport packaging and remove the plastic cover. Wait
until the device has adjusted to the ambient temperature.
2. Remove the foamed plastic packaging from the SPECORD® sample compartment.
The packaging contains the items included in the standard equipment (see above)
and the installation disks for the WinASPECT
®
optional module you ordered.
3. Install the software on the PC (
→
Refer to WinASPECT
®
User's Manual).
4. Insert the deuterium lamp following the instructions given in Section 8.3 Changing
5. Connect the SPECORD® (
→
see 6.1 Connectors and display elements p. 21) to
the PC (free COM port) using the supplied connecting cable.
6. Connect one power cable each to the SPECORD® and the PC. Connect the power
cables of the SPECORD® and the PC to power outlets that are connected to the
same phase of the line power supply (multiple socket outlet, if possible).
Page 24
SPECORD
®
200/205/210/250 User's Manual – Edition 04/2005
Analytik Jena AG