Maintenance and care
62
Edition 05/2014
PlasmaQuant PQ 9000
−
Pour the prepared coolant into the tank until the MAX level mark is reached.
−
Switch on the recirculating chiller and observe the level indicator. The level drops when
the pump is operating.
−
Continue to slowly fill the tank with the coolant until the level stabilizes slightly below
the MAX mark.
−
Remove the funnel and seal the filling opening with the screw cap.
−
Confirm the closing step in the wizard.
3. Wait for the message in the wizard indicating that the coolant has been changed. Exit the
wizard.