Fault List
The Fault List screen is a read-only screen that allows the
user to display the cause of the highest priority failure. The
screen indicates the number of faults present in the system
and a message detailing the highest priority fault present.
Note that some faults can result in multiple displayed failures
due to the high number of internal tests occurring. As faults
are corrected, they are immediately cleared.
Faults are not stored; therefore, they are immediately
removed if power is cycled. If the problem causing the faults
still exists, however, faults will be displayed again after
power is re-applied and a period of time elapses during
which the diagnostic system re-detects them. The
exception to this rule is the calibration failure. When a
calibration fails, no corrupt data is stored. Therefore, the
system continues to function normally on the data that was
present before the calibration was attempted.
After 30 minutes or if power to the transmitter is cycled, the
failure for calibration will be cleared until calibration is
attempted again. If the problem still exists, the calibration
failure will re-occur. Press ENTER to initiate view of the
highest priority failure. The display will automatically return
to normal after a few seconds.
PID Timer
This function sets a timer to monitor the amount of time the
PID controller remains at 0% or 100%. This function only
appears if the PID controller is enabled. If the timer is set to
0000, the feature is effectively disabled. If the timer value is
set to any number other zero, a FAIL condition will occur if
the PID controller remains at 0% or 100% for the timer value.
If one of the relays are set to FAIL mode, this failure
condition can be signaled by a changing relay contact.
Press ENTER to initiate user entry mode, and the entire
value will flash. Use the UP arrow key to modify desired
value; range of value is 0-9999 seconds. Press ENTER to
store the new value.
Sim Out
The Sim Out function allows the user to simulate the oxygen
level of the instrument in the user selected display range.
The user enters a ppm value directly onto the screen, and
the output responds as if it were actually receiving the signal
from the sensor. This allows the user to check the function
of attached monitoring equipment during set-up or