ATI
Q46D-ODO Optical D.O. System
Part 4 – Electrical Installation
20
O&M Manual
Rev-J (8/17)
4.3
Relay Connection
Three SPDT relays are provided on the power supply board. None of the relay
contacts are powered. The user must supply the proper power to the contacts.
For applications that require the same switched operating voltage as the Q46
(115 or 230 VAC), power may be jumped from the power input terminals at
TB7.
Relay wiring is connected at
TB4, TB5, and TB6
as shown below. Note that the
relay contact markings are shown in the NORMAL mode. Programming a relay
for “Failsafe” operation reverses the NO and NC positions in this diagram (Figure
11).
Figure 11 - Relay Contacts