Part 5 – Configuration
40
O&M Manual
Rev-J (8/17)
*Set 4 mA #2
*Set 20 mA #2
[temp/D.O.]
These functions set the second 4 mA and 20 mA current
loop output points for the transmitter. The output may be set
to track temperature (default), PPM oxygen, or % Saturation.
The values stored for the 4 mA point may be higher or lower
than the value stored for the 20 mA point.
The entry value is limited to a value between 0 and 50 °C if it
is set for temperature. The 4 mA and the 20 mA point must
be at least 20 units away from each other. Press ENTER to
initiate user entry mode, and the value will flash. Use arrow
keys to modify value. Press ENTER to store the new value.
NOTE: I
f the temperature units are changed between °C and
°F (see Temp Units in this section), the default settings for
this output will be stored (present data is not converted.)
*Set 4 mA #3
*Set 20 mA #3
[temp/D.O.]
OPTIONAL.
These functions set the optional third 4 mA and
20 mA current loop output points for the analyzer. The
output may be set to track temperature (default) or D.O. The
values stored for the 4 mA point may be higher or lower than
the value stored for the 20 mA point.
The entry value is limited to a value between 0 and 55 °C if it
is set for temperature. The 4 mA and the 20 mA point must
be at least 20 units away from each other. Press ENTER to
initiate user entry mode, and the value will flash. Use arrow
keys to modify value. Press ENTER to store the new value.
*A Setpoint
This function establishes the oxygen trip point for relay A.
The entry value is limited to a value within the range
specified in “Set Range”. Use the LEFT arrow key to select
the first digit to be modified. Then use the UP and LEFT
arrow keys to select the desired numerical value. Press
ENTER to store the new value.
*A Hysteresis
This function establishes the hysteresis, or “deadband”, for
Relay A. Hysteresis is most often used to control relay
chattering; however, it may also be used in control schemes
to separate the ON/OFF trip points of the relay. Press
ENTER to initiate user entry mode, and the value will flash.