ATI
Q45C2 2E Conductivity System
Part 4 – Configuration
O & M Manual
Revision K (01/07)
- 23 -
Cord-grips used for sealing the cable should be snugly tightened
after electrical connections have been made to prevent moisture
incursion. When stripping cables, leave adequate length for
connections in the transmitter enclosure as shown below. The
standard 15 ft. sensor cable normally supplied with the system is
already stripped and ready for wiring. This cable can be cut to a
shorter length if desired to remove extra cable in a given
installation. Do not cut the cable so short as to make installation
and removal of the sensor difficult.
.
Figure 3-5 Sensor Cable Preparation