Analox 50 – Carbon Dioxide Analyser – User Manual
Document Ref: A50-811-19 – September 2014
Page 21
7. TECHNICAL
SPECIFICATIONS
Carbon Dioxide
Range 0.1-5%
Response Time
60 Seconds to T90
Optional Oxygen Range
0.1 to 25%
Operating Temperature
-5 to +40°C
Temperature Effect
<0.1%FS/°C
Sensor Type - Carbon Dioxide
Analox Infra Red Detector, featuring
microprocessor based compensation for
temperature effects and IR source ageing
Sensor Type – Oxygen
Galvanic Electrochemical Cell
Optional Display
4 digit Liquid Crystal Display
Optional 4 to 20mA Output (or 0-1V)
Gas Concentration Output
(Max load 150
Ω
)
Alarms
2/3 x Carbon Dioxide Visual Indicators
1 x System Fault Indicator
1 x Status Indicator
1 x Optional Oxygen Visual Indicator
Common Audible Alarm - Horn
Relays Internal:
2 x Optional Independent Relays, Volt-
Free single pole Changeover, rated
50VAC/DC, 1 Amp, (only one relay if
oxygen sensor specified).
or
External:
2 x Optional Independent Relays, Volt-
Free single pole Changeover, rated
240VAC/ 28VDC, 2 Amps, (only one relay
if oxygen sensor specified).
Analox has a policy of continuous improvement and we reserve the right
to upgrade or change specifications without prior notice.
The Analox 50
TM
conforms to DIN 6653-2:2004.