Analox 50 – Carbon Dioxide Analyser – User Manual
Document Ref: A50-811-19 – September 2014
Page 12
4. INSTALLATION
4.1. Wall Mounting
The power socket that the equipment plugs into must be positioned so
that it is close to the equipment and in easy reach of the operator.
If the power supply for the Ax 50
TM
is not positioned so that it is close to
the equipment and in easy reach of the operator, a local means of
double pole isolation must be provided. Isolation can be achieved either
by means of a non-locking plug or a double pole switch of suitable rating.
A label must be placed by the local disconnection point for the Ax 50
TM
stating
"REMOVE PLUG TO DISCONNECT Ax 50
TM
"
The Ax 50
TM
should be mounted onto a wall using the mounting lugs on
the unit. Rawl plugs and screws are provided for this purpose. It is not
necessary to dismantle the Ax 50
TM
main unit in any way prior to
installation. You need to ensure the mains plug fused at 5amp is in easy
reach of a power socket.
Attach the main unit to the wall 450mm from the floor as close to the
valves and manifolds as possible. An 8 metre 8 core cable is pre-wired
to the Ax 50
TM
and has a Remote Alarm Repeater pre-attached. Run the
Remote Alarm Repeater cable to the outlet door, safely securing the
cable with cable clips.
A drilling template is included with this booklet.