6
Settings
Applicable actuator
Description
Range
Defaut
settings
Min limit #6 (% or °)
Actuator #6 (Right
stander)
Minimum setting of right stander actuator (Standing module fully
deployed) minimum angle between the seat plate and the base
Notes 6 et 7
-70° to 0°
-68
Max limit #6 (% or °)
Actuator #6 (Right
stander)
Maximum setting of right stander actuator (Standing module
completely closed)
Notes 6 et 7
-70° to 0°
0
•
Note 1 :
Not applicable on actuators for very heavy-duty seat tilt (450 lb to 550 lb).
•
Note 2 :
If the chair has the power standing option, this setting should never be other than "0", otherwise
the standing function will not work.
•
Note 3 :
Setting the height limit can be determined with the following "rule of three": Height limit value
needed in % = (Height limit value in inches x 100)/11 inches (maximum height range).
E.g. Height limit value needed is 5.5 inches
→
(5.5 X100)/11 = 50%.
Therefore, the height limit value of
Max limit #3
is 50%
•
Note 4 :
Not applicable on actuators for heavy-duty seat elevate (300 lb to 400 lb).
•
Note 5 :
If the chair has the power standing option, minimum setting of center mount legrest is 21°. It is
possible to change this value, but watch for leg hyperextention when in standing mode.
•
Note 6
: Minimum and maximum limits for the left and right stander actuators must be set at the same
value.
•
Note 7
: It is strongly recommended not to set this limit lower that -68° (example -70°) because the legrest
mechanism may collide with the base of the chair.
•
Note 8
: This setting should never be other than "0" as this would prevent the seat tilt and seat elevate
from functionning.
•
Note 9
:
WARNING!
If the chair has the power standing option, it is not recommended to change this
value because the backrest will not completely open to its vertical position which makes it dangerous for
the user who may risk being thrown forward.
Settings under the “Actuator Setup” submenu