Settings
Applicable actuator
Description
Range
Defaut
settings
Min limit #1 (% or °)
Actuator #1 (Tilt)
Note 1
Minimum setting for tilt position (Angle between seat plate and chair
base).
Note 2
0° to 46°
0
Max limit #1 (% or °)
Actuator #1 (Tilt)
Note 1
Maximum setting for tilt position (Full tilt) max angle between seat
plate and base.
0° to 46°
46
Min limit #2 (% or °) Actuator #2 (Recline)
Minimum setting for backrest recline position (Angle between seat
plate and chair base).
91° to 160°
91
Max limit #2 (% or °) Actuator #2 (Recline)
Maximum setting for backrest recline position (Fully reclined)
maximum angle between seat plate and base.
Note 9
91° to 160°
160
Min limit #3 (% or °)
Actuator #3 (Elevate)
N
ote 4
Minimum setting for seat elevate position (Percentage of the range
of elevation).
Note 2
0% to
100%
0
Max limit #3 (% or °)
Actuator #3 (Elevate)
N
ote 4
Maximum setting for seat elevate position (Fully elevated)
percentage of range.
Note 3
0% to
100%
100
Min limit #4 (% or °)
Actuator #4 (Center
mount)
Minimum setting of center mount legrest position (Legrest fully
elevated) minimum angle between legrest and seat plate line.
Note 5
92° to 5°
5
Max limit #4 (% or °)
Actuator #4 (Center
mount)
Maximum setting of center mount legrest position (Legrest fully
lowered).
92° to 5°
92
Min limit #5 (% or °)
Actuator #5 (Left
legrest)
Minimum setting of left legrest position (Legrest fully elevated)
minimum angle between legrest and seat plate line.
67° to 11°
11
Max limit #5 (% or °)
Actuator #5 (Left
legrest)
Maximum setting of left legrest position (Legrest fully lowered).
67° to 11°
67
Min limit #6 (% or °)
Actuator #6 (Right
legrest)
Minimum setting of right legrest position (Legrest fully elevated)
minimum angle between legrest and seat plate line.
67° to 11°
11
Max limit #6 (% or °)
Actuator #6 (Right
legrest)
Maximum setting of right legrest position (Legrest fully lowered).
67° to 11°
67
Min limit #5 (% or °)
Actuator #5 (Left
stander)
Minimum setting of left stander actuator (Standing module fully
deployed) minimum angle between the seat plate and the base
Notes 6 et 7
-70° to 0°
-68
Max limit #5 (% or °)
Actuator #5 (Left
stander)
Maximum setting of left stander actuator (Standing module
completely closed)
Notes 6 et 7
-70° to 0°
0