12
•
Note 1:
Resetting actuators using the first special boot-up mode
For custom limits: Note all settings beforehand. Replace them with values shown in the Default Settings
column of the General submenu table (Section 2.1.2.1) and then proceed to a position reset of actuator(s).
Restore the settings to their original values before returning the chair to the user.
•
Note 2:
Resetting stander actuators using the first special boot-up mode
For the standing function, the center mount legrest actuator minimum limit and that of the stander
actuators are not set to the minimum physical limit of these actuators. Since the value shown in the
Default Settings column of the General submenu table (Section 2.1.2.1) is not the minimum physical limit
of stander actuators, the values must be replaced as follows:
→
Once these values are set, procced with a position reset of the center mount legrest actuator or of the stander
actuators through the first special boot-up mode.
Important!
Prior to modifying settings, take note of the original values. These values will be restored before
returning the chair to the user.
3.4
Second special boot-up mode
Once the Second special boot-up mode is accessed, a position reset of actuators can be done by moving the
targeted actuators from the minimum position to the maximum position using the keypad (SWKEYPAD). Retract all
actuators to neutral position before exiting this mode and return to normal mode by turning chair off and back on.
To execute a position reset, refer to the Features column in the table below.
Table for the Second special boot-up mode
Number of green LED
flashes at start-up on
the e-bx
Mode 2 indications
What becomes
disabled
Applicable
actuators
Features
2
•
Red and green
LEDs on the
pushbutton
keypad
(SWKEYPAD)
constantly flash.
•
Power options
are not visible on
the joystick’s
display in
“Seating” mode
•
All position limits
•
All error alarms
that may occur
•
All automatic
actuator
functions and
memory
positions.
All actuators
See
Note 1.
For stander
actuators
See
Note 2
•
Allows a minimum and maximum position reset of
actuators.
•
Works only with pushbutton keypad (SWKEYPAD).
•
Only actuators connected to channels A to E can be
activated. If needed, connect channel F actuator to
channel E to move it in this mode. The left pushbutton
activates channel A, the second from the left activates
channel B and so on.
•
All automatic functions such as stander actuator
management during deployment and memory position
recall are disabled. All actuators become independent.
•
The chair goes into drive lock-out and can no longer
move.