78
MU Series (Windows)
User’s Manual
Process Menu Tools
Extended Depth of Focus (EDF)
5. The final menu for the EDF process is the alignment options menu. Since there can be small variations in
X- and Y- dimensions even when moving only the focusing knob (based on how a rack and pinion system is
machined when moving forcus can cause variations), the auto alignment process attempts to correct for these
variations to avoid a blurred image, or a misaligned image when stacking the images together.
As before, the menu has several notes as to this menu’s purpose and function, shown below.
Setting it to “None” will assume that the images are perfectly align and will not perform correction processes.
Setting it to “Shift + Scale” will attempt to translate directionally as well as increase/decrease in size for best
alignment.
Setting it to “Shift + Scale + Rotate” will do the same as above, however, will also rotate the image if it was
rotationally skewed during focus translation of the stage.
When finished, press “Finish” to begin the software processing of the image stack.