3
switches, socket-outlets in the fixed installation, and
equipment for industrial use and some other equipment,
for example, stationary motors with permanent
connection to the fixed installation.
CENELEC Directives
The instruments conform to CENELEC Low-voltage
directive 2006/95/EC and Electromagnetic compatibility
directive 2004/108/EC
X�
Warning: Read Before Using
• To avoid possible electrical shock or personal injury,
follow these instructions and use the Meter only as
specified in this manual.
• Do not use the Meter or test leads if they appear
damaged, or if the Meter is not operating properly.
If in doubt, have the Meter serviced.
• Always use the proper function and range for
measurements.
• Before rotating the function range selection switch,
disconnect test probe from circuit under test.
• Verify the Meter’s operation by measuring on a
known voltage source.
• Do not apply more than the rated voltage, as
marked on the Meter, between the test probe or
between any test probe and earth ground.
• Use the Meter with caution for voltages above 30
Vac rms, 42 Vac peak, or 60 Vdc. These voltages pose
electrical shock hazards.
• Disconnect circuit power and discharge all high-
voltage capacitors before testing resistance.
• Do not use the Meter around explosive gas or vapor.
• When using the test leads, keep your fingers behind
the finger guards.
• Remove test leads from the Meter before opening
the Meter case or battery door.
• if the meter is used in a manner not specified in
the users manual, the protection provided by the
equipment may be impaired
www.
.com
1.800.561.8187