Measuring AC and DC Current
Press SELECT button for switching to DC current
measurement function.
X�
To avoid personal injury or damage to the Meter:
1. Do not attempt to make an in-circuit current
measurement when the open-circuit potential to
earth ground exceeding 600 V.
2. Switch to proper function and range for your
measurement.
3. Do not place the test probe in parallel with a circuit
when the test leads are connected to the current
terminals.
4. Connect the test leads to the correct input A/mA μA
current terminal and to the circuit before powering
the circuit under test.
5. After measurement, switching OFF the circuit’s
power before removing test leads from the circuit.
or
for
for
For current range from 8-10A, do not measure current
for more than one (1) minute. Wait for 10 minutes
before taking another measurement.