EMC GUIDANCE & MANUFACTURER’S DECLARATION
33
D-0115690-C (OM001)
– Otowave 102 Operating Manual
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic
environment
– guidance
Voltage
dips,
short
interruptions
and
voltage variations on
power supply input lines
IEC 61000-4-11
<5% U
T
(>95% dip in U
T
) for
0.5 cycle
40% U
T
(60% dip in U
T
) for
5 cycles
70% U
T
(30% dip in U
T
) for
25 cycles
<5% U
T
(>95% dip in U
T
) for 5
sec
Not applicable
Not applicable
Power frequency (50/60
Hz) magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location
in
a
typical
commercial
or
hospital
environment.
NOTE U
T
is the a.c. mains voltage prior to the application of the test level