TAKING MEASUREMENTS
11
D-0115690-C (OM001)
– Otowave 102 Operating Manual
If reflex test(s) were carried out these results will now be displayed:
The display shows:
The frequency and level of the reflex stimulus
“PASS” if a reflex was found, else “x” (No Response)
A plot of compliance against time
If the reflex test was performed at more than one frequency use the
▲ and ▼ keys to view the results for the
other frequencies.
If
the Otowave 102 was set to test for a reflex at all levels of the stimulus press ► to view an additional display
following the reflex graphs. This shows a summary of the levels and frequencies at which a reflex was detected.
The dash symbol “-” is shown if a reflex tone was not presented at the level indicated.
REFLEX SUMMARY
dB
100
X
-
90
X
80
X
70
X
X
X
Hz 500 1k 2k 4k
Press ◄ to return and view the tympanogram, reflex results or to repeat the test. When satisfied with the
results press ►.
The message “Saving as last test” will be displayed and the results will be saved in the “last test” memory. The
results will remain available until a new test is started, even if the Otowave is turned off.
If both ears were chosen for test the entire sequence will now be repeated for the right ear:
TESTING RIGHT EAR
INSERT PROBE
Cancel Skip
Press ► to skip testing of the right ear and display the PROCESS RESULTS menu. Press ◄ to cancel and
return to the ear selection menu. In both cases the left ear results are retained and may be viewed as the LAST
TEST.
Otherwise insert the probe; the right ear test will then proceed as described above.