PERFORMING DAILY CHECKS
24
D-0014930-A
– Otowave 102-C Operating Manual
11.
PERFORMING DAILY CHECKS
The operation of the Otowave 102-C should be checked daily using the 4 in 1 test cavity supplied with the
instrument.
Select the DAILY CHECK option in the main menu:
DAILY CHECK
INSERT PROBE
Cancel
Wait until “Open” is displayed.
Insert the probe, without an ear tip, into the hole at the 2ml end of the test cavity. Make sure that the probe is
pushed in fully and is held tight against the stop. The probe must be square to the end of the test cavity.
The display should show the volume of the test cavity to within ± 0.1ml.
DAILY CHECK
Volume: 2.0 ml
Cancel
Remove the probe and repeat the test with the three remaining test cavities. The display should show the
volume of the 0.2ml and 0.5ml test
cavities to within ± 0.1ml. The 5.0ml test cavity should be within ± 0.25ml.
When the checks have been completed press
◄ to return to main menu.