TAKING MEASUREMENTS
15
D-0014930-A
– Otowave 102-C Operating Manual
TESTING LEFT EAR
Obtaining ear seal
Low :
High :
Cancel
The number of bars shown indicates the robustness of the seal. The probe should be adjusted in the ear until
two or more bars are shown for Low & High.
5.5.
ERROR MESSAGES
The following error messages may be seen during the test sequence.
MESSAGE
DISPLAYED
INDICATOR
STATUS
LIKELY CAUSE(S)
WITHDRAW PROBE
Yellow Flashing
The probe has been moved during measurement.
Re-insert the probe to repeat the test.
Volume outside range
WITHDRAW PROBE
Yellow Flashing
The ear canal volume is above the 5ml. This
message can also occur when the probe is not
properly inserted into the ear.
Blocked ear
WITHDRAW PROBE
Green Flashing
The ear canal volume is below 0.1ml. Check that the
probe is not blocked and correctly inserted into the
ear.
INSERT PROBE
Yellow Flashing
The seal was lost. Reinsert the probe to repeat the
test.