User Manual for
AML Oceanographic’s
Plus
X
27
Output Format Commands
Command
Description
Requires
DISPLAY SCAN
Displays current scan options.
SET SCAN NOBAT
Turns the battery channel off.
SET SCAN BAT
Turns the battery channel on.
SET SCAN NODENSITY Turns the calculated density channel off.
C
•X
, T
•X
, and P
•X
,
CT
•X
, and P
•X
, or
SV
•X
, T
•X
, and P
•X
SET SCAN DENSITY
Turns the calculated density channel on.
C
•X
, T
•X
, and P
•X
,
CT
•X
, and P
•X
, or
SV
•X
, T
•X
, and P
•X
SET SCAN NOSALINITY Turns the calculated salinity channel off.
C
•X
, T
•X
, and P
•X
,
CT
•X
, and P
•X
, or
SV
•X
, T
•X
, and P
•X
SET SCAN SALINITY
Turns the calculated salinity channel on.
C
•X
, T
•X
, and P
•X
,
CT
•X
, and P
•X
, or
SV
•X
, T
•X
, and P
•X
SET SCAN NOSV
Turns the calculated salinity channel off, removing it from
the instrument output scans. Current salinity display status
is viewable using DISPLAY SCAN.
C
•X
, T
•X
, and P
•X
or
CT
•X
and P
•X
SET SCAN SV
Turns the calculated sound velocity channel on, allowing it
to be present in instrument output scans. Current salinity
display status is viewable using DISPLAY SCAN. This is
only available when C,P and T sensors are attached.
C
•X
, T
•X
, and P
•X
or
CT
•X
and P
•X
SET SCAN TIME
Enables displaying time in data scan.
SET SCAN NOTIME
Disables time from being displayed in data scan.
SET SCAN DATE
Enables displaying date in data scan.
SET SCAN NODATE
Disables date from being displayed in data scan.
DISPLAY STARTUP
Displays the power up output settings.
SET STARTUP
PROMPT
Sets the instrument to wait for user commands on power up.
SET STARTUP SCAN
Sets the instrument to output one scan on power up and
then wait for a user command.
SET STARTUP
MONITOR
Sets the instrument to start monitoring data on power up.
SET STARTUP
NOHEADER
Disables the instrument identification header output on
power up.
SET STARTUP
HEADER
Enables the instrument identification header output on
power up.
Logging Commands
Command
Description
Requires
SET SCAN LOGGING
Enables simultaneous logging & real-time output. If real-time
logging in air is desired, set instrument conductivity
threshold and sound velocity and pressure increments to
zero.
SET SCAN
NOLOGGING
Disables simultaneous real-time logging.
SET TIMEOUT nn
nn is time in minutes from 0 to 30. Enters logging mode after
the specified time interval has passed in which the
instrument has been idle. Power the unit off, then on, to exit