Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
The equipment or system is intended for use in the electromagnetic environment specified below. The customer or
the user of the equipment or system should assure that it is used in such an environment.
Immunity Test
IEC 60601 Test Level
Compliance Level
Electromagnetic
Environment –
Guidance
Electrostatic
Discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood,
concrete or ceramic
tile. If floors are covered
with synthetic material,
the relative humidity
should be at least 30%.
Electrical Fast
Transient/Burst
IEC 61000-4-4
±2 kV for power
supply lines
±1 kV for input/
output lines
±1 kV for power supply
lines
±0.250 kV for input/
output lines
Mains power quality
should be that of a
typical commercial or
hospital environment.
Surge
IEC 61000-4-5
±1 kV line(s) to
line(s)
±2 kV line(s) to earth
±1 kV line(s) to line(s)
±2 kV line(s) to earth
Mains power quality
should be that of a
typical commercial or
hospital environment.
Voltage Dips, Short
Interruptions and
Voltage Variations on
Power Supply Input
Lines IEC 61000-4-11
<5 % UT
(>95 % dip in UT)
for 0,5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT)
for 5 sec
<5 % UT
(>95 % dip in UT)
for 0,5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT)
for 5 sec
Mains power quality
should be that of a
typical commercial or
hospital environment.
If the user of the
equipment or
system requires
continued operation
during power mains
interruptions, it is
recommended that the
equipment or system
be powered from an
uninterruptible power
supply or a battery.
Power Frequency
(50/60 Hz) Magnetic
Field IEC 61000-4-8
3 A / m
Not Applicable
Power frequency
magnetic fields
should be at levels
characteristic of a
typical commercial or
hospital environment.
NOTE: UT is the A.C. mains voltage prior to application of the test level.
Electromagnetic Compliance Data for iCE Series