Installation: Amico iCE LED Light
!
CAUTION -
Installation and removal for service or inspection must be carried out by a hospital technician or a
person with similar qualifications who has been trained by Amico Clinical Solutions Corp.
• Never perform an installation or removal of the LED light alone. Always have a second person to assist with the
installation or removal.
• Always disconnect the spring arm system from the power supply when installing or removing the LED light.
Failure to do so may result in electric shock.
• Always make sure the spring arm is at the highest position when removing the light. Failure to do so will cause
the spring arm to spring upwards, and may cause serious injury.
• The LED light may separate from the spring arm if the key is not properly in place. This may cause serious
injury. Always make sure that the key is installed properly.
• Failure to comply with load specifications may cause serious or fatal injuries. Ensure that the LED light load is
within the specifications.
Installing the Lighthead
Normal System/LCH
1. Ensure the unit is disconnected from the power supply, with no possible accidental reconnection.
2. Apply approved grease to the male end of the lighthead and locking key before installation, to ensure smooth
movement.
3. Ensure the safety plug is removed prior to lighthead installation.
PROJECT MANAGER : ___________________
ENGINEERING : __________________________
PRODUCTION : __________________________
55 East Wilmot St.
Richmond Hill, Ontario
L4B 1A4, CANADA
Tel: (905)764-0800
Fax: (905)764-0862
DESCRIPTION:
PART NO:
SHEET:
OF
DRAWING NO:
PROJECTION
DATE:
DATE:
CHECKED BY:
RC
DRAWN BY:
SCALE:
C