Troubleshooting
Should problems arise with the use of the Amico Series iCE LED Lights, review the following chart. Find the fault
and complete the recommended solution. If the fault is not found and or/the solution does not correct the problem
contact Amico Clinical Solutions Corp.
Fault
Recommended Solution
The emergency red light indicator is “On”
• Switch “Off” the emergency switch to cancel Emergency By-Pass mode
• If one or two LED Pods stop working, the fuse may be blown and it may
need to be replaced by a qualified technician
A button on the main mylar controller has
stopped functioning
• The mylar controller may need to be replaced
A button on the wall mylar controller has
stopped functioning
• The wall mylar controller may need to be replaced
Lighthead is drifting
• Tighten up the exposed screws on the yokes or the central axis using a
3/8" flat head screwdriver.
Glass is dirty
• Follow “Operation: Cleaning the Amico iCE LED Light”
An LED is not functioning
• The pod needs to be replaced by a qualified technician
Maintenance: OASYS Spring Arm
Annual Inspection (to be completed by qualified technicians):
6. All labels are attached and are legible.
7. All spring arm components are not deformed.
8. All moving parts are free from squeaking and are not loose.
9. Rotational stops are working correctly.
10. All components are free from collision damage.
11. Welds are free from cracks.
12. Load, parallel, brake and vertical adjustments are conducted to ensure the spring arm is performing correctly.
Refer to section, Installation: Adjustments, pg. 33-37.