40
Amico Clinical Solutions Corp.
Installation: Wiring Diagrams
Negatives
Segment 1
Segment 2
Communication Line
for Wall Control 1
and Lighthead 1
Communication Line
for Wall Control 2
and Lighthead 2
Dual LED Configuration with Wall Control
1. Insert the following positive lines into SEGMENT 1 as shown in the image below.
a. Lighthead 1
b. Wall Control 1
c. Power Supply 1
4. Insert the following positive lines into SEGMENT 2 as shown in the image below.
a. Lighthead 2
b. Wall Control 2
c. Power Supply 2
4. Match the communication cables of the WALL CONTROLS and the LIGHTHEADS in the terminal blocks labeled
C1 and C2.
5. Insert the remaining negative lines into the terminal block labeled “-VE”.
2.
3.
4.