9
Amico Accessories Inc.
1. First Time Initializing
:
Plug cart into power outlet – Advanced Interface should turn on (may take up to 30 seconds). After initializing the
cart, a “No Home“ symbol ( ) will appear in the top right hand corner of the Advanced Interface. This means the
user must set the cart's "Home" (lowest) position.
2. Setting Home Position:
1. Hold the Down arrow to lower the cart's work surface until the collision symbol ( ) appears.
2. Release the Down arrow and reverse direction to raise work surface to desired position.
3. Setting Time and Date:
1. Tap on “Guest“ in top left corner of interface
(Figure 1)
.
2. Select “Admin“ from drop down list and login using password: AMICO
(Figure 2)
.
3. Tap on “Settings“ icon.
4. Select “Admin Setup“ (Only available in the ADMIN profile).
5. Select “Set Clock“.
6. Using the arrow icons set the time and date in the desired format (note: date and time formats must first be
selected on right of interface). Tap Save to finish.
Figure 1
Settings
Notes
4:50 pm
07/10/2014
Timer
10h:50m
GUEST
Top
Middle
Bottom
Figure 2
4:50 pm
07/10/2014
10h:50m
Guest
USER NAME
PASSWORD
LOG IN
NEW USER
CANCEL
Q
W
E
R
T
Y
U
I
O
P
A
S
D
F
G
H
J
K
L
Enter
Z
X
C
V
B
N
M
.
abc
Bk
123
123
#+=
#+=
Section 3: Software Initialization
Installation and Adjustment Instructions
Powered Cart with Advanced Interface (Automatic Height Adjustment)
Quick Setup
Medical Facility Responsibility
WARNING:
THIS QUICK-GUIDE IS NOT A REPLACEMENT FOR THE FULL MANUAL. IT IS THE RESPONSIBILITY OF THE
END USER TO ENSURE ALL ASPECTS OF THE INSTALLATION ARE COVERED BY FOLLOWING THE FULL MANUAL.
NOTE:
INSPECT CART INTEGRITY PRIOR TO BEGINNING SETUP.