www.amico.com
25
Appendix D
WIRING DIAGRAM:
LCD DISPLAY BOARD - LOCAL SENSOR
B4
B3
B2
B1
SE
TUP
CHANGE UP
CHANGE DOWN
SELEC
T/MUTE
AC Supply
115 to 220 V
AC
DC P
ower Cable
:
Connect to
Annunciator Module
G
N
L
L - Live
N - Neutral
G - Ground
Fuse (1 AMPw)
Ground
Toggle S
wit
ch
OXYGEN
# 22 Gauge twisted
pair shielded cable
6”-8” [0.1m-0.2m] supplied
CAUTION:
To protect from static electricity,
ensure to discharge body static before
installing the Medical Gas Alarm and
sensors
NOTE:
Do not ground the shield drain wire at
sensor or inside alarm panel back box
Sensor
Module
Black
Red
6" - 8" [0.1m - 0.2 m]
#22 gauge stranded,
shielded and twisted
pair cable supplied
Orange (15 VDC)
Blue (5 VDC)
Black (Ground)