24
Amico Corporation
Appendix C
WIRING DIAGRAM:
LCD DISPLAY BOARD - ALARM BUZZER
B4
B3
B2
B1
SE
TUP
CHANGE UP
CHANGE DOWN
SELEC
T/MUTE
AC Supply
115 to 220 V
AC
DC P
ower Cable
:
Connect to
Annunciator Module
G
N
L
L - Live
N - Neutral
G - Ground
Fuse (1 AMPw)
Ground
Toggle S
wit
ch
NOTE:
Relays on the annunciator are fail
safe for Version 3.1 or newer.
Relays are not fail safe for Versions
3.0 or older.
NO COM NC 12V GND
NC COM NO 12V GND
NO COM NC 12V GND
Optional: To Amico Manifold Universal
Remote Buzzer
NOTE:
Relays on the annunciator are fail
safe for Version 3.1 or newer.
Relays are not fail safe for Versions
3.0 or older.
NO COM NC 12V GND
NC COM NO 12V GND
NO COM NC 12V GND
NOTE:
AMICO RECOMMENDS MAX. 50 FT. TO POWER UP BUZZER FROM ALARM PANEL TO POWER UP THE BUZZER.
MORE THAN 50 FT., A A3P-POWER-V4 IS REQUIRED TO SUPPLY VOLTAGE FOR THE ALARM BUZZER.
Optional: Abnormal Alarm
Generic Alarm Dry-
Contact Fault Indicator
(to Building Management
System/Master)
Orange (15 VDC)
Blue (5 VDC)
Black (Ground)