E c o n o m y M a n i f o l d
P a g e : 1 2
TROUBLE SHOOTING
This section is intended to serve as a general guide for identifying the potential functional problems
which occur in operation of Amico manifolds.
When an asterisk (*) appears beside the CORRECTIVE ACTION, the recommended procedure is
to replace the whole unit in question, with a substitute unit, until repairs are completed on the original
equipment.
Components removed for maintenance, must be serviced, repaired and tested, only by personnel
qualified to work on equipment used in medical service. Only original manufacturer’s parts, as
supplied by Amico, may be used in the maintenance of Amico manifolds.
S Y M P T O M
C A U S E C O R R E C T I V E
A C T I O N
Operating Pressure Regulator Faults
Gas leakage around oper-
ating pressure regulator
body cap.
Venting at intermediate
relief valve.
Pipeline not at desired
pressure.
Required gas flow not
available.
Low pressure relief valve
venting.
1
. Tighten
cap.
2
. * Replace regulator with
substitute unit and
change diaphragm.
1.
*Replace regulator with
substitute unit and
repair.
1.
Set line regulator to value
speciafied on (page 8)
Line Regulator.
1.
Set line regulator to value
specified on (page 8)
Line Regulator.
1.
Set line regulator to value
speciafied on (page 8)
Line Regulator.
2.
Replace the relief valve
with a new one set in
accordance with ( page
9) Pressure Relief Valve.
Line Pressure Regulator Faults
a.
Loose
cap.
b.
Diaphragm
leak.
a.
Over pressure due to
creeping or faulty
regulation by operting
pressure regulator.
a.
Line regulator not set
correctly.
a.
Line regulator not set
correctly.
a.
Line regulator set at too
high a delivery pressure.
b.
Relief valve set at too
low a pressure.