32 Amico Corporation
WIRING DIAGRAM:
LCD DISPLAY MODULE - REMOTE SENSOR
Note:
For multiple sensors, a
multi-conductor twisted
pair cable can be used.
OXYGEN
# 22 Gauge twisted
pair shielded cable
6”-8” [0.1m-0.2m] supplied
Junction Box
(by others)
Sensor
Module
DISS
Oxygen
Pipeline
Remote location alarm
AC Supply
115 to 220 V
AC
DC P
ower Cable
:
Connect to
Annunciator Module
G
N
L
L - Live
N - Neutral
G - Ground
Fuse (1 AMPw)
Gr
ound
To
ggle S
wit
ch
COM
B4
et
u
M /
tc
el
e
S
N
W
O
D
eg
na
h
C
P
U
eg
na
h
C
p
U -
te
S
B3
B2
B1
+12V
NO
NC
GRND
Reset
Button
S 8
S 7
S 6
S 5
S 4
S 3
S 2
S 1
NOTE:
For multiple sensors, a
multi-conductor #22 Gauge
stranded, shielded and
twisted pair cable ONLY
must be used.
Black
Red
Appendix E
6" - 8" [0.1m - 0.2 m]
#22 gauge stranded,
shielded and twisted
pair cable supplied
#22 gauge stranded,
shielded twisted pair
cable only must be
used, up to a distance of
2000 ft [609.6 m]. In the
presence of any electrical,
magnetic radio, wireless
or other interference,
cable installation MUST
be placed in a metallic
conduit
Marrette the sensor
cable in a junction box
(supplied by others) to
the installation cable
(supplied by others)
Содержание A3P-SENS-L-GAS