STEP 9: LCD DISPLAY SETUP
i. Press the SETUP button (B1) and press the SELECT button (B4)
ii. Volume control: 90, 80, 70, 60 – press CHANGE UP/DOWN to change noise level
iii. Press SELECT for LCD brightness and press CHANGE UP/DOWN to change LCD brightness
iv. Press the SELECT button (B4)
* The following should be displayed.
SENSOR: 1
SENSOR TYPE: show the gas (example: NITROUS OXIDE)
UNITS: PSI (to change to kPA, Bar, inHg, mmHg: press button CHANGE DOWN/UP then press SELECT)
LOW ALARM: 40 (to change: press button CHANGE DOWN/UP then press SELECT)
HIGH ALARM: 60 (to change: press button CHANGE DOWN/UP then press SELECT)
CURRENT PRESSURE: displays exact pressure for the line
CURRENT OFF SET: to re-calibrate pressure reading press button CHANGE DOWN/UP then press SELECT
v. Repeat steps until all sensors are scanned and data is saved. Next it will show SETUP COMPLETE.
*All gases should now be displayed. If any errors occur, repeat the steps above.
NOTE
: Hold the “MUTE” button for twenty (20) seconds to display current low and high set points.
NOTE
: Press the Setup button (B1) in order to make corrections/go back.
STEP 10: CLOSING THE FRAME/MODULE ASSEMBLY
i. Close the frame panel by tightening the screws found on the frame panel to the back box. Ensure that the
screws are securely fastened to keep the LCD Alarm closed.
ii. Carefully place the front frame over the frame panel. Screw in the screws that were removed in
Step 5,
part ii
. The alarm shall now be ready for use.
Installation Guide
CAUTION:
To protect from static electricity, ensure to discharge body static before installing the Medical Gas
alarm and sensors.
A m i c o M i c r o p r o c e s s o r B a s e d A l a r m
P a g e : 4
USER RESPONSIBILITY
The information contained in this Installation and Operation Maintenance
Manual, pertains only to the ALERT-2 microprocessor based digital alarm.
This product will perform to conformity with the descriptions contained in this
manual, when assembled, operated, maintained and serviced in accordance
with the installation instructions provided.
The alarm
must
be checked periodically. Parts that are broken, missing,
worn, distorted or contaminated,
must
be replaced immediately. Should such
repair or replacement become necessary, please contact Amico Corporation
or their distributors.
All alarms should not be repaired, or altered without prior written or verbal
approval of Amico Corporation or it’s distributors. Failure to comply will void
all warranty on the alarm.
Statements in this manual preceded by the words
WARNING
,
CAUTION
,
DANGER
and
NOTE
are of special significance. Please read these sections
carefully.
WARNING:
denotes steps which can prevent injury.
CAUTION:
denotes steps which can prevent damage to equipment.
DANGER:
denotes steps which can prevent electrical shock to equipment
or to prevent serious injury and/or death.