Specifications and Disclaimer
37
Specifications and Disclaimer
Specifications:
210BR Standard ranges:
0 – 1%, 0 – 5%, 0 – 10%, 0 –25%
Sensitivity: 0.5% of full scale
Repeatability: +/- 1% of full scale at constant temperature
Operating temperature: -4
C to 46
C (25
F - 115
F); -29
C to 46
C (-20
F - 115
F) with optional heater
Humidity: < 85%, non-condensing
Operational conditions: Pollution degree 2, Installation category I I.
Drift: +/- 1% of full scale in 4 weeks at constant temperature (dependent on sensor).
Expected cell life: 6 months to 2 years (dependent on sensor).
Response times (upscale):
90% of full scale in less than 30 sec (dependent on sensor).
Outputs: 1-5V or 4-20mA isolated.
Alarm contacts: 117VAC @ 5A, or 28VDC @ 5A, resistive.
Power requirements: 12-24VDC, <80mA, or 117VAC <1W; power increases to ~25W with optional
heater.
Absolute Maximum Power voltage 28V DC or 128VAC 60Hz.
Overall dimensions: 15” w x 10” h x 6¼” d.
Mounting hole dimensions: 2.8” w x 9.3” h.
Weight 15 lbs.
117VAC version is to be installed only in installation (overvoltage) category I or II.