AMI Analyzer Manual
Maintenance and troubleshooting
28
Maintenance and troubleshooting
Maintenance:
The model 210BR is virtually maintenance free other than for periodic calibration and occasional sensor
replacement.
Periodic Calibration:
The analyzer should be calibrated about once every month to obtain the best accuracy. The sensor typically
declines in sensitivity by about 1% per month, so a monthly calibration is usually satisfactory. Use in a
particularly aggressive environment may degrade the sensor faster: in this case calibrate more often.
You can see the calibration history using the user interface as described in the previous section of this
manual.
Sensor Replacement:
This should be done based on the Span Factor feature, rather than as a response to a dead sensor. See the
chart below for recommended sensor replacement.
Sensor
Part
number
Description
Warrantied life
P-2
4SEN03
Percent oxygen 0-25% - inert gas
9 months
P-3
4SEN04
Percent oxygen – 0-25% CO2 background
6 months
P-4
4SEN08
Percent oxygen – 0-100% - inert gas
6 months
P-5
4SEN19
Percent oxygen – 0-25% CO2 with 500ppm
H2S
6 months
Table 1. AMI percent sensor types
Содержание 210BR
Страница 1: ...08 31 2018 Made in the USA AMI Oxygen Analyzer Manual Model 210BR Costa Mesa CA ...
Страница 2: ...AMI Analyzer Manual Contents i ...
Страница 16: ...AMI Analyzer Manual Installation Procedure 13 ...
Страница 24: ...AMI Analyzer Manual Analyzer Description 21 Drawings Figure 7 Outline Drawing Figure 8 Flow schematic ...
Страница 25: ...AMI Analyzer Manual Analyzer Description 22 Figure 9 Control drawing AC unit Figure 10 Control drawing DC unit ...