AMI Analyzer Manual
Installation Procedure
19
If it doesn’t, something is wrong. If the sensor reads very low, check it on air, and if that reading is also very
low, replace the sensor (you shouldn’t have to do this with a new sensor!) If the reading is too high, it is
possible that your span gas is contaminated, or you have a leak, or you have made some kind of error. One
thing to note is that if the analyzer temperature is very different from the sensor temperature when you put
a new sensor in, the reading will be quite wrong until the sensor has had a chance to equilibrate.
i)
If so, adjust the analyzer span until it reads the span gas value.
Press the SPAN button. The SPAN flag should appear on the LCD. Press the UP or DOWN arrow buttons
until the reading is correct. After a few seconds the SPAN flag will go out and the analyzer will accept the
new calibration.
j)
Let it go back to normal operation (the “SPAN” flag goes out on the LCD display), then press
the UP arrow and note the number displayed (the “Calibration factor”).
You can use this to get an idea of the remaining cell life in the future. The analyzer also stores the
calibration values in memory so you can view the history later. When the sensor is new it will have a
Calibration factor (also called Span factor) of about 450. As it gets used up, this number will have to
increase when you span the analyzer until it gets up to around 950 or so (the Calibration factor doesn’t
increase by itself, it only does so when you span the analyzer). You will probably see the span factor remain
fairly constant, then suddenly change. When it does so it means that you will need a new sensor soon. The
sensor behaves rather like a NiCd battery – it is stable until suddenly it starts to fail.
k) Rotate the valve back to the Sample position.
The analyzer is now set up and calibrated. It is normally good practice to give it a day or two and recheck
the calibration.
Содержание 210BR
Страница 1: ...08 31 2018 Made in the USA AMI Oxygen Analyzer Manual Model 210BR Costa Mesa CA ...
Страница 2: ...AMI Analyzer Manual Contents i ...
Страница 16: ...AMI Analyzer Manual Installation Procedure 13 ...
Страница 24: ...AMI Analyzer Manual Analyzer Description 21 Drawings Figure 7 Outline Drawing Figure 8 Flow schematic ...
Страница 25: ...AMI Analyzer Manual Analyzer Description 22 Figure 9 Control drawing AC unit Figure 10 Control drawing DC unit ...