AMI Analyzer Manual
Maintenance and troubleshooting
••••
52
3. Verify that the output connections are not shorted all the way back to the recording device. Disconnect
the wires from the analyzer and use an ohmmeter to check for shorts or opens.
4. Verify that the unit is set to the correct output. Remove the output board cover and verify that the
jumpers are in the correct positions (on the left for current output, on the right for voltage output). If
not, place them correctly.
No output alarm indication
1. Verify the alarm set points are correct - press the appropriate switch on the front panel, and check the
displayed reading on the LCD for correct setting.
2. Verify that the connections on the terminal block are properly stripped and correct.
3. Verify that the alarms are configured correctly, using a PC or similar communication device.
4. Verify the alarm delay time with the PC.
5. Verify that the output connections are not shorted all the way back to the recording device. Disconnect
the wires from the analyzer and use an ohmmeter to check for shorts or opens.
Incorrect readings
1. Verify that there are no leaks in the system.
2. Verify that the span gas bottle is correctly marked by comparing its reading when the analyzer has been
spanned on air to what it actually says.
3. If spanning on air, verify that the air source is free of water vapor (humid air will contain about 3% less
oxygen than expected, depending on temperature), and that bottle air does actually contain 20.9%
oxygen. Manufactured air often does not!
Analyzer refuses to accept front panel settings
1. Using the RS-232 interface, verify that the security is set the way you want it.
Still no correct operation
1.
Call AMI at 714 848 5533, and ask for Technical
assistance.
2.
Or contact us by email at sales@
AMIO
2.com.