AMI Analyzer Manual
Communications
••••
47
Diagnostic Functions
The diagnostic functions 0,1,2,4,10,11,12,13,14,15, and 16 are supported.
Note that each counter will count up to 65535, but won’t go any higher. They can be reset to zero with the
10 command.
Function
Command
(without CRC)
Action
Notes
0
11 08 00 00
Echo message
Returns the exact same message.
1
11 08 00 01
Restart communication
Restarts from a previous 4 command
2
11 08 00 02
Return error byte
Returns same as holding register 23
4
11 08 00 04
Listen only mode
Stops the analyzer from responding to anything.
10
11 08 00 0A
Clear all diagnostic counters
Clear each of the diagnostic counters to zero
11
11 08 00 0B
Total message count
The total number of messages seen by the
analyzer
12
11 08 00 0C
CRC error count
Number of CRC errors seen by the analyzer
13
11 08 00 0D
Exception count
Number of invalid Modbus commands
14
11 08 00 0E
Number of slave messages
Number of messages the analyzer has returned
15
11 08 00 0F
Number of no responses
Number of messages addressed to the analyzer
that it did not respond to
16
11 08 00 10
Number of NAK responses
Number of messages with incorrect parameters
(such as invalid registers or out-of-bounds values)
seen by the analyzer