Communications
••••
30
Communications
Basics
The 201RSP supports both a USB link to a local PC, and RS485 over a network. Only one of these can be
active at a time, and the unit automatically turns off the RS485 if it detects a computer plugged into the USB
connection. Therefore the PC connection over the USB has priority.
The USB connection supports the proprietary AMI protocol, and AMI provides a program on a CD that
interfaces with the analyzer. Alternatively it is possible for someone familiar with communications on a PC
to directly use the protocol.
The RS485 port supports ModBus RTU at 9600 baud.
USB communication:
The USB port is located on the cover panel in the explosion proof enclosure, and is provided as a type B USB
connector. It uses a standard cable such as are often used to connect PCs to printers.
Through the USB interface you can operate the analyzer from your PC, and you can set up its internal
parameters to your liking. The USB communication runs at 57600 baud.
The RS-485 Modbus™ connection is disabled whenever the USB connector is plugged in (at both ends).
•
We recommend that you use the AMI communication program for reading from and setting up the
analyzer.
•
Use the Version 3 or later program (identifiable by having a blue background to its main screen). It
supports the new analyzer features mentioned below.