User Manual | 39
Calibration Screen
Pressing the Page key to display the calibration information page.
The calibration page is used to calibrate the instrument and shows
information about the most recent run. Name is the gas species to be
measured.
Conc.
is the concentration (or blend) of the calibration gas
connected to the instrument.
Area
is the peak area of the impurity de-
tected during the last analysis run for each impurity type.
Rfactor
is the response factor calculated by the system on the most re-
cent calibration. Rfactor is used to calculate the concentration of impuri-
ties.
RT
is retention time of the impurity. This is the time correspond-
ing to the top of each peak calculated from the start of the most recent
anlaysis run.
Calibration
Screen
Pressing page again will exit the page screens.
It is recommended to use the Ametek’s Trace Analytical MGB1000
Micro Gas Blender to blend the calibration gas to the desired level
when calibrating the ta5000R