User Manual | 17
Sample Gas
Inlet Fitting
1/16-inch VICI compression fitting
Flow Rate
20 - 100 sccm minimum
Inlet pressure stability
± 2%, UHP regulator required
Vent pressure
Atmospheric pressure vent is optimal,
±
0.5
psig
maximum
Actuator Gas
Inlet Fitting
1/16-inch VICI compression fitting (Uses
Sample Inlet)
Inlet Pressure Range
70 to 150 psig (4.8 to 10.2 bar)
Gas Type
CDA, dewpoint , -65° C
Calibration Gas
Inlet Fitting
1/16-inch VICI compression
fitting(Uses
Sample
Inlet)
Cylinder Concentration
H
2
, CO, Balance nitrogen, depends
on
level
of
detection
Cylinder Volume
100 cubic feet minimum (2800L)
Chassis
Dimensions
7” (18cm) H x 16.8” (43cm) W x 26.5”
(67cm)
D
Weight
35
lb.
(15.9
kg)
Power
100 - 120 VAC, 50/60 Hz,;
200 - 240 VAC, 50/60 Hz