38 | Model ta5000R Gas Purity Monitor
Diagnostic Print Key
The Print key appears when setting the clock. Pressing the Print key from
this page will print a complete report of all system settings, calibration
settings, temperatures and status and the chromatogram from the last
run of the analyzer. This is a system diagnostic tool which can be used to
troubleshoot problems with the analyzer.
Saving changes into Non-Volatile RAM
After changes have been made to any of these fields the system LED will
turn red. This indicates that there is a system error or system state change.
Use the
Page
key to move to the System Settings page. Any change made
will be shown under the System Errors line. Use the arrow keys to move
the cursor in front of the change you wish to accept and press the key
under the
Accept
menu label.
Next, press the
Save
key to save the accepted changes into Non-Volatile
RAM. The instrument must be in “Idle” mode to accept the changes.
After changes have been made to any of these fields the system LED
will turn red. This indicates that there is a system error or system
state change.
Temperatures and Status Page
Pressing the Page key twice will display the Temperatures and Status
Page. This page shows status information about the critical heated zones
and status of the detector. No user input is allowed on this page in nor-
mal user mode.
Image of Temperatures and Status Screen
!
CAUTION
Temperatures & Status
Idle Mode
12:34:25
Page
Meas
Set
RGD Bed = 265 # 265
RGD Column = 128 # 130
RGD Lamp = On
RGD Vout = 36187
RGD Vsig = 1978.4
------------------------- Pressure --------------------------
Carrier = 20.3
Содержание TRACE ANALYTICAL ta5000R
Страница 14: ...14 Model ta5000R Gas Purity Monitor This page left blank...
Страница 23: ...User Manual 23 Display screen...
Страница 48: ...48 Model ta5000R Gas Purity Monitor Dimensional Drawings ta5oooR 3 D View...
Страница 49: ...User Manual 49 22 115 6 92 ta5oooR Side View...
Страница 50: ...50 Model ta5000R Gas Purity Monitor 24 070 16 70 ta5oooR Top View...