8-8 | Thermox WDG-V
RM
Alarm and Warning Messages
AlArMs
AlArM
DesCriPTion
Possible Causes
Cell T/C Failure
Thermocouple is not connected or is open
Bad T/C, Bad connection
Cold Junction Compensator
Failure
The PCB temperature sensor has failed
Bad Electronics
Cell Temperature Control
A critical overtemperature condition has
occurred
Shorted, open, or Bad
T/C or Failed Temperature
Control Circuit.
Unit must be reset to clear
this alarm
Cell Over Temp
Cell temperature is 4 degrees or higher
above the setpoint
Bad Thermocouple, Failed
temperature control circuit
Cell Under Temp
Cell temperature is 4 degrees or lower
below the setpoint
Faulty heater, faulty heater
connection, Faulty T/C,
Faulty T/C connection,
Low Power.
Only after warm-up is
complete
Cell Temp Rise Failure
Cell heater failed to heat the cell
Bad Cell furnace heater,
Bad heater connection,
Shorted T/C, Bad T/C con-
nection.
Unit must be reset to clear
this alarm
Over Temp Relay Tripped
The cell over-temperature relay has tripped Open Thermocouple, Over
Temperature condition,
Failed Temperature control
circuit.
Must cycle power to reset
the relay
RTD Failure
The Box Temp RTD Failed
Open or Shorted RTD
Low Sample Flow
Low sample flow detected
Loss of sample flow.
Active only after warm-up
time expires
Cell Failure (Open)
Cell resistance exceeds the normal limit
Bad Cell, Open Cell Con-
nection
Cell Mv Mismatch
The cell mv measurement does not match
the real time cell meas
Bad A/D circuit (Bad elec-
tronics)
Analog Output 1 Error
The analog output measurement does not
match the setting
Open 4-20mA loop or
failed output or bad read-
back circuit
Cell T/C Measurement Mis-
match
The Redundant temperature measurement
does not match the real time meas
Bad T/C or circuit
Содержание Thermox WDG-VRM
Страница 1: ...Thermox WDG VRM User Manual Thermox 150 Freeport Road Pittsburgh PA 15238 9000 166 VE REV H...
Страница 13: ...System Overview 1 3 Figure 1 1 WDG VRM Schematic...
Страница 24: ...2 4 Thermox WDG VRM This page intentionally left blank...
Страница 27: ...Installation 3 3 Figure 3 1 WDG VRM front view Figure 3 2 WDG VRM mounting dimensions...
Страница 34: ...3 10 Thermox WDG VRM Figure 3 7a Customer I O Interconnecting WDG VRM...
Страница 35: ...Installation 3 11 B DETAIL B SCALE 6 4 Figure 3 7c Customer I O Interconnecting WDG VRM...
Страница 44: ...3 20 Thermox WDG VRM This page intentionally left blank...
Страница 103: ...Customer Modbus Map 10 1 Customer Modbus Map...