Configuration
|
5-5
Setting the Alarms
The WDG-V
RM
has five independent, normally open alarms. The relays can be set
to energize or deenergize on alarm.
Alarms: Process, Service, Fault and Data Valid
There are three (3) process relay alarms, (1) fault alarm and (1) data alarm.
Process alarms
are activated when the process reading levels, based on what the
user has defined in the setup menus, have been exceeded.
Service alarms
activate when there is a hardware malfunction and the analyzer is
not operating correctly.
Data Valid alarms
activate when there is a hardware malfunction or when the output
may not be reflective of the process, for example during Warm up, Calibration or
Diagnostic modes.
Process alarms are configurable for:
Function (O2)
High/Low limit
Use Left/Right Arrow to navigate to Alarm Configuration.
Figure 5-5. Configuration Menu Screen