4-22 | Thermox CG1000 Oxygen Analyzer
Relay Configure
All Series 2000 control unit relays are Form A—normally open contacts.
You can use the software to set whether the alarms should close on an
alarm condition (energize) or open on an alarm condition (de-energize).
Configuring the relay
1. From the
Alarm
menu, select
Relay Configure
.
2. Select
De-energize on Alarm
or
Energize on Alarm
.
If you select
Energize on Alarm
, the following appears:
Relays are set to
ENERGIZE on Alarm
If you select
De-Energize on Alarm
, the following appears:
Relays are set to
DE-ENERGIZE on Alarm
Select Function
Select the type of reading that Alarm 3 should track.
If tracking the start of a calibration or verification, the alarm will deacti-
vate once the calibration or verification has been completed, including
anyr ecovery duration you have specified.
Alarm 4 is always set to track oxygen and therefore is not included in
the Select Function menu.
Defining the select function
•
From the
Alarm
menu, choose
Select Function
.
•
Select
Alarm 3= O2/Flow
,
Alarm 3= In Cal
, or
Alarm 3= In Verify
,
depending on whether Alarm 3 should be associated with oxygen
levels, flow rate, or with the start of a calibration or verification.
The CG1000 automatically exits this menu after you make your selection.