Load Connection and Sensing
Local Sensing
31
Local Sensing
Sensing of the output voltage is available from both the rear panel and the front panel
output connectors. Default local sensing regulates the voltage at the power supply
output terminal. Use remote sensing (see “Remote Sensing” on page 32) when the
voltage needs to be regulated at the load rather than at the power supply output
terminals.
Without sense line connections, the supply regulates the voltage at the output
terminals of the power supply.
See Figure 1.1 Front Panel Controls, p. 13, and Figure 1.3 Rear Panel., p. 14, for
sense terminal locations on the front and rear panels.