Local Operation
User Diagnostics
65
Table 3.4
Quick Diagnostics
Symptom
Check
Further Checks and Corrections
No output and the
display is blank.
Is input voltage in the specified range?
Connect to appropriate voltage source.
Power switch ON?
Turn on power.
Internal circuit.
See your service technician.
No output but the
display turns on.
SW1 switch settings correct? (For when
using analog programming.)
Output Voltage and Current Limit” on
.
OVP LED turned on?
See
“Using Over Voltage Protection
.
Front panel S/D LED turned on?
See
“Using the Shutdown Function” on
.
OTP LED turned on?
Current limit set to zero?
See
“Standard Operation” on page 51
.
Voltage control set to zero?
See
“Standard Operation” on page 51
.
REM LED turned on?
If using remote analog control, check your
analog programming source. (See
Output Voltage and Current Limit” on
.) If not, refer to your digital
interface manual.
Is front panel ACF LED turned on?
Connect unit to AC supply in specified
range.
Internal circuit.
See your service technician.
No output and
display blinks
dimly.
Internal circuit.
See your service technician.
Output not
adjustable.
Is unit in current limit mode? (Red
Current Mode LED turned on).
Turn current knob clockwise to increase
current limit. Reduce load if current is at
maximum. See
Automatic Crossover” on page 51
.
Is unit in remote mode? (Green REM
LED turned on).
If using remote analog control, check your
analog programming source. (See
Output Voltage and Current Limit” on
.) If not, refer to your digital
interface manual.
Is unit at maximum voltage or current
limit?
Reduce load for lower voltage or current
requirement.
Содержание Sorensen XHR 100-10
Страница 2: ......
Страница 4: ...This page intentionally left blank ii ...
Страница 6: ...This page intentionally left blank iv ...
Страница 8: ...This page intentionally left blank vi ...
Страница 12: ...List of Figures x Operating Manual for XHR Series Power Supply ...
Страница 14: ...List of Tables xii Operating Manual for XHR Series Power Supply ...
Страница 47: ...Installation Load Connection 47 Figure 2 9 Multiple Loads with Remote Sensing ...