16060-101 Rev. M
45
Guidance Tables
(
continued)
Table 9
– Guidance and Manufacturer’s Declaration
Electromagnetic Immunity
Immunity to Proximity Fields from RF Wireless Communications Equipment
Guidance and Manufacturer’s Declaration - Electronic Immunity
The Reichert 7CR is intended for use in the electromagnetic environment as specified below related
to proximity fields from RF wireless communications equipment.
Immunity
Test
IEC 60601Test Level
Compliance
Level
Electromagnetic
Environment
-Guidance-
Radiated RF
IEC 61000-4-3
Test
Frequency
(MHz)
Band (MHz)
Service (MHz)
Modulation
Maximum
Power (W)
Distance
(m)
Immunity
Test Level
(V/m)
Compliance
Level
d = 6/E √P
where
d = Minimum
separation
distance in
meters
E = Immunity
test level in
V/m
P = Maximum
power in
Watts (W)
385
380-390
TETRA 400
Pulse Modulation
18 Hz
1,8
0,3
27
27 V/m at
0,3 m
450
430-470
GMR 460, FRS
460
FM ±5 kHz devia-
tion 1 kHs sine
2
0,3
28
28 V/m at
0,3 m
710
704-787
LTE Band 13, 17
Pulse Modulation
217 Hz
0,2
0,3
9
9 V/m at 0,3 m
745
780
810
800-960
GSM 800/900,
TETRA 800,
iDEN 820, CDMA
850, LTE Band 1,
3, 4, 25; UMTS
Pulse Modulation
18 Hz
2
0,3
28
28 V/m at
0,3 m
870
930
1720
1700-1990
GSM 1800; CD-
MA 1900; GSM
1900; DECT; LTE
Band 1, 3, 4, 25;
UMTS
Pulse Modulation
217 Hz
2
0,3
28
28 V/m at
0,3 m
1845
1970
2450
2400-2570
Bluetooth WLAN,
802.11 b/g/n,
RFID 2450, LTE
Band 7
Pulse Modulation
217 Hz
2
0,3
28
28 V/m at
0,3 m
5240
5100-5800 WLAN 802.11 a/n Pulse Modulation
217 Hz
0,2
0,3
9
9 V/m at 0,3 m
5500
5785
Содержание Reichert 7CR